Liquistation CSF33 Operation
Endress+Hauser 45
10.2.3 Hold settings
Menu/Setup/General settings/Hold settings
Function Options Info
Settings automatic Hold
Hold release time 0...600 s
Factory setting
0 s
The hold is maintained for the duration of the
delay time when you switch to the measuring
mode.
Setup menu Selection
• Disabled
• Enabled
Factory setting
Disabled
Diagnostics menu Decide whether a hold should be output at the
current output when the particular menu is
opened.
Calibration active Factory setting
Enabled
If a device-specific hold is enabled, any cleaning that was previously started is
stopped. You can only start a manual cleaning if a hold is active. The hold has no
influence on the sampling.
10.2.4 Logbooks
Logbooks record the following events:
• Calibration/adjustment events
• Operator events
• Diagnostic events
• Programming events
You define how the logbooks should store the data.
In addition, you can also define individual data logbooks .
1. Assign the logbook name.
2. Select the measured value to be recorded.
3. Set the recording rate (Scan time).
You can set the recording rate individually for every data logbook.
Further information on the logbooks: .
Menu/Setup/General settings/Logbooks
Function Options Info
Logbook ident Customized text, 16
characters
Part of the file name when exporting a logbook
Event logbook Selection
• Off
• Ring buffer
• Fill up buffer
Factory setting
Ring buffer
All diagnostic messages are recorded
Ring buffer
If the memory is full, the most recent entry
automatically overwrites the oldest entry.
Fill up buffer
If the memory is full, there is an overflow,i. e.
you cannot store any new values. The controller
displays a corresponding diagnostic message.
The memory then has to be cleared manually.