Liquistation CSF33 Operation
Endress+Hauser 47
Menu/Setup/General settings/Logbooks
Function Options Info
Start/stop
simultaneously
Action Appears if you have created more than one data
logbook. With one mouse click, you can start or
stop recording all the data logbooks.
Logbook name The name of this submenu is based on the name
of the logbook and only appears once you have
created a logbook.
This menu appears several times if you have several data logbooks.
Source of data Read only This is for information purposes only. If you
want to record another value, delete this logbook
and create a new data logbook.
Measured value
Log time left
Event logbook =
Fill up buffer
Read only Displays the days, hours and minutes remaining
until the logbook is full.
Log size
Event logbook =
Fill up buffer
Read only Displays the number of entries remaining until
the logbook is full.
Logbook name Customized text, 20
characters
You can change the name here again.
Scan time 0:00:01 to 1:00:00
Factory setting
0:01:00
As above
Minimum time interval between two entries
Format: H:MM:SS
Data logbook Selection
• Ring buffer
• Fill up buffer
Factory setting
Ring buffer
Ring buffer
If the memory is full, the most recent entry
automatically overwrites the oldest entry.
Fill up buffer
If the memory is full, there is an overflow, i. e. no
new values can be saved. The controller displays
a corresponding diagnostic message. The
memory then has to be cleared manually.
Overflow warnings
Event logbook =
Fill up buffer
Selection
• Off
• On
Factory setting
Off
‣
Decide whether you want to receive a
diagnostic message if the fill buffer of the
relevant logbook overflows.