Liquistation CSF33 Operation
Endress+Hauser 75
Menu/Setup/Sampling programs/Setup program/New/Basic
Function Options Info
Bottle change mode Selection
Number of samples
Time
External signal
Factory setting
External signal
The bottle can be changed after a specific
number of samples, after a time or by an
external signal.
Having selected Bottle change mode Number of samples
Samples per bottle 1 ... 9999
Factory setting
1
Set the number of samples.
Having selected Bottle change mode Time
Time interval 00-00:02 ... 31-00:00
DD-HH:MM
Factory setting
0001:00 DDHH:MM
Set the time (days, hours and minutes) after
which the system should change to the next
bottle.
Multiple bottles 0 ... 23
The configuration options
depend on the current
number of bottles
Factory setting
0
Multiple bottles:
"Simultaneous" transfer of two samples to
separate bottles.
Having selected Bottle change mode External signal
Bottle chg. sig. input Selection
No bottle change input
configured
Binary input Sx
Factory setting
No bottle change input
configured
The bottle change input can be configured
under Inputs .
A binary input must be configured for this
function. The sampling input can be
configured under "Inputs".
Multiple bottles 0 ... 23
The configuration options
depend on the current
number of bottles
Factory setting
0
Multiple bottles:
"Simultaneous" transfer of two samples to
separate bottles.
Start condition Selection
Immediate
Date/time
Factory setting
Immediate
The sampling program can be started either
immediately or at a specific, configurable
time.
Having selected Start condition Date/time
Start date 01.01.2000 ... 31.12.2099
Factory setting
DD.MM.YYYY
Set the start date for the sampling program.
The format depends on the option
configured under general settings.
Start time 00:00:00 ... 23:59:59
Factory setting
HH:MM:SS (24h)
Set the time when the sampling program is
started. The format depends on the option
configured under general settings.
Stop condition Selection
Program end
Continuous
Factory setting
Program end
Program end
The device stops sampling automatically
once it has run through the set program.
Continuous
The device runs through the set program
continuously in an infinite loop. Do not
forget to empty the bottles.