EasyManua.ls Logo

Endress+Hauser Liquistation CSF33 - Page 76

Endress+Hauser Liquistation CSF33
180 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation Liquistation CSF33
76 Endress+Hauser
Menu/Setup/Sampling programs/Setup program/New/Basic
Function Options Info
Assignment bin. output Selection
No binary output
configured
Binary output Sx
Factory setting
No binary output
configured
Assignment of the binary output to the
program cycle.
Inputs Settings for the inputs can be made as
described in the "Inputs" section.
Settings with a Basic program and external signal
Settings with the Basic program type via an external signal with 1 bottle
Sampling mode = External signal
Menu/Setup/Sampling programs/Setup program/New/Basic
Function Options Info
Sampling volume input 10 to 1000 ml
Factory setting
100 ml
Enter the sample volume.
Sampling signal input Selection
No sampling input
configured
Factory setting
No sampling input
configured
Select the input for the sampling signal. The
fieldbus must be configured for this
function.
The sampling input can be configured under
Inputs .
Bottle change mode Selection
Number of samples
Time
Number of samples
Factory setting
Number of samples
The bottle can be changed after a specific
number of samples, after a time or by an
external signal.
Having selected Bottle change mode Number of samples
Samples per bottle 1 ... 9999
Factory setting
1
Set the number of samples.
Having selected Bottle change mode Time
Time interval 00-00:02 ... 31-00:00
DD-HH:MM
Factory setting
0001:00 DDHH:MM
Set the time (days, hours and minutes) after
which the system should change to the next
bottle.
Bottle synchronization Selection
None
1. bottle change time
1. Time of change +
bottle number
Factory setting
None
None
The sampling and bottle change times are
not synchronized.
1. bottle change time
Sampling starts with the first bottle. Set the
synchronization time.
1. Time of change + bottle number
Each bottle is assigned to a specific fill time.
Set the synchronization time and the
weekday.

Table of Contents

Other manuals for Endress+Hauser Liquistation CSF33

Related product manuals