Liquistation CSF33 Operation
Endress+Hauser 77
Menu/Setup/Sampling programs/Setup program/New/Basic
Function Options Info
Start condition Selection
• Immediate
• Date/time
Factory setting
Immediate
The sampling program can be started either
immediately or at a specific, configurable
time.
Having selected Start condition Immediate
Sample at start Selection
• Yes
• No
Factory setting
Yes
Yes
The first sample is taken when the program
is started.
No
The system waits for the interval to elapse
before the first sample is taken.
Having selected Start condition Date/time
Start date 01.01.2000 ... 31.12.2099
Factory setting
DD.MM.YYYY
Set the start date for the sampling program.
The format depends on the option
configured under general settings.
Start time 00:00:00 ... 23:59:59
Factory setting
HH:MM:SS (24h)
Set the time when the sampling program is
started. The format depends on the option
configured under general settings.
Stop condition Selection
• Program end
• Continuous
Factory setting
Program end
Program end
The device stops sampling automatically
once it has run through the set program.
Continuous
The device runs through the set program
continuously in an infinite loop. Do not
forget to empty the bottles.
Assignment bin. output Selection
• No binary output
configured
• Binary output Sx
Factory setting
No binary output
configured
Assignment of the binary output to the
program cycle.
Inputs Settings for the inputs can be made as
described in the "Inputs" section.
Settings with the Basic program type via an external signal with multiple bottles
Sampling mode = External signal
Menu/Setup/Sampling programs/Setup program/New/Basic
Function Options Info
Sampling volume input 10 to 1000 ml
Factory setting
100 ml
Enter the sample volume.
Sampling signal input Selection
No sampling input
configured
Factory setting
No sampling input
configured
Select the input for the sampling signal. The
fieldbus must be configured for this
function.
The sampling input can be configured under
Inputs .