Operation Liquistation CSF33
78 Endress+Hauser
Menu/Setup/Sampling programs/Setup program/New/Basic
Function Options Info
Bottle change mode Selection
Number of samples
Time
Number of samples
Factory setting
Number of samples
The bottle can be changed after a specific
number of samples, after a time or by an
external signal.
Having selected Bottle change mode Number of samples
Samples per bottle 1 ... 9999
Factory setting
1
Set the number of samples.
Having selected Bottle change mode Time
Time interval 00-00:02 ... 31-00:00
DD-HH:MM
Factory setting
0001:00 DDHH:MM
Set the time (days, hours and minutes) after
which the system should change to the next
bottle.
Having selected Bottle change mode External signal
Bottle chg. sig. input Selection
No bottle change input
configured
Binary input Sx
Factory setting
No bottle change input
configured
The bottle change input can be configured
under Inputs .
A binary input must be configured for this
function. The sampling input can be
configured under "Inputs".
Multiple bottles 0 ... 23
The configuration options
depend on the current
number of bottles
Factory setting
0
Multiple bottles:
"Simultaneous" transfer of two samples to
separate bottles.
Start condition Selection
Immediate
Date/time
Factory setting
Immediate
The sampling program can be started either
immediately or at a specific, configurable
time.
Having selected Start condition Immediate
Sample at start Selection
Yes
No
Factory setting
Yes
Yes
The first sample is taken when the program
is started.
No
The system waits for the interval to elapse
before the first sample is taken.
Having selected Start condition Date/time
Start date 01.01.2000 ... 31.12.2099
Factory setting
DD.MM.YYYY
Set the start date for the sampling program.
The format depends on the option
configured under general settings.
Start time 00:00:00 ... 23:59:59
Factory setting
HH:MM:SS (24h)
Set the time when the sampling program is
started. The format depends on the option
configured under general settings.