Operation Liquistation CSF33
80 Endress+Hauser
Menu/Setup/Sampling programs
Function Options Info
Start condition Selection
• Immediate
• Date/time
• Volume
Factory setting
Immediate
The sampling program can be started either
immediately, at a specific, configurable time,
or when a certain totalized flow is reached.
Having selected Start condition Date/time
Start time 00:00:00 ... 23:59:59
Factory setting
HH:MM:SS (24h)
Set the time when the sampling program is
started. The format depends on the option
configured under general settings.
Start date 01.01.2000 ... 31.12.2099
Factory setting
DD.MM.YYYY
Set the start date for the sampling program.
The format depends on the option
configured under general settings.
Having selected Start condition Volume
Start volume input Selection
• No flow input
configured
• Binary input S:x
• Current input S:x
Factory setting
No flow input configured
Select the start volume input. The binary
input or the current input must be
configured for this function. Only the inputs
configured for flow measurement are
displayed.
Start flow sum 1000 to 9,999,000 m
3
Factory setting
10,000 m
3
Set the start volume.
Stop condition Selection
• Program end
• Continuous
• Date/time
Factory setting
Program end
Program end
The device stops sampling automatically
once it has run through the set program.
Continuous
The device runs through the set program
continuously in an infinite loop. Do not
forget to empty the bottles.
Date/time
The device stops the set program at a
specific time.
Having selected Start condition Date/time
Stop date 01.01.2000 ... 31.12.2099
Factory setting
DD.MM.YYYY
Set the stop date for the sampling program.
The format depends on the option
configured under general settings.
Stop time 00:00:00 ... 23:59:59
Factory setting
HH:MM:SS (24h)
Set the time when the sampling program is
stopped. The format depends on the option
configured under general settings.
Setup subprogram
New
Programpart Use a distinct name for your subprogram.
The program name can be up to 16
characters long.