Operation Liquistation CSF33
84 Endress+Hauser
Programming example: Program switchover
The following example explains how to program a program change in the Standard
program.
Flow-paced sampling
Daily average sample
From Monday to Friday
Time-paced sampling
In a composite container
On Saturday and Sunday
Menu/Setup/Sampling programs/Setup program/New
Function Entry for sample program
Standard
Program name TDVT 6+20
Bottle configuration 6x + 1x
Bottle volume left 3000 ml
Bottle volume right 20000 ml
Start condition Date/time
Start date DD.MM.YYYY e.g. 03.06.2010
Start date HH:MM:SS e.g. 12:00:00
Stop condition Continuous
Setup subprogram
Programpart MO FR
"SAVE" saves the "MO FR" subprogram
Setup subprogram
Programpart SAT SUN
"SAVE" saves the "SAT SUN" subprogram
"ESC" the main program is displayed
Bottle assignment Statical bottle assignment
Programpart: MO FR
Edit
Sampling mode Flow paced VTCV
Flow measurement Current input S:1
Sampling interval 1000 m
3
Sampling volume 80 ml
Bottle change mode Time
Time interval 04:00
Multiple bottles 0
Enable subprogram Multiple date
Start condition No delay
Activity time 24:00
Repeating date
Repetition mode Days of week
Monday Yes
Tuesday Yes