Operation Liquistation CSF33
84 Endress+Hauser
Programming example: Program switchover
The following example explains how to program a program change in the Standard
program.
Flow-paced sampling
• Daily average sample
• From Monday to Friday
Time-paced sampling
• In a composite container
• On Saturday and Sunday
Menu/Setup/Sampling programs/Setup program/New
Function Entry for sample program
Standard
Program name TDVT 6+20
Bottle configuration 6x + 1x
Bottle volume left 3000 ml
Bottle volume right 20000 ml
Start condition Date/time
Start date DD.MM.YYYY e.g. 03.06.2010
Start date HH:MM:SS e.g. 12:00:00
Stop condition Continuous
Setup subprogram
Programpart MO ‐ FR
"SAVE" ‐ saves the "MO ‐ FR" subprogram
Setup subprogram
Programpart SAT ‐ SUN
"SAVE" ‐ saves the "SAT ‐ SUN" subprogram
"ESC" ‐ the main program is displayed
Bottle assignment Statical bottle assignment
Programpart: MO ‐ FR
Edit
Sampling mode Flow paced VTCV
Flow measurement Current input S:1
Sampling interval 1000 m
3
Sampling volume 80 ml
Bottle change mode Time
Time interval 04:00
Multiple bottles 0
Enable subprogram Multiple date
Start condition No delay
Activity time 24:00
Repeating date
Repetition mode Days of week
Monday Yes
Tuesday Yes