Liquistation CSF33 Operation
Endress+Hauser 83
Menu/Setup/Sampling programs
Function Options Info
Sample at disable Selection
• Yes
• No
Factory setting
No
Specify whether a sample should be taken
when the subprogram is disabled. For
example, with intervals, a sample is taken at
the end of every activation interval.
New bottle at disable Selection
• Yes
• No
Factory setting
Yes
Bottle synchronization Selection
• None
• 1. bottle change time
• 1. Time of change +
bottle number
• External BC sync input
Factory setting
None
Specific bottles can be assigned specific
filling times with the bottle synchronization
function. For example, bottle 1 is to be filled
from midnight to 2 a.m., bottle 2 from 2
a.m. to 4 a.m. etc..
None
The sampling and bottle change times are
not synchronized.
1. bottle change time
Sampling starts with the first bottle. The
change to the next bottle is synchronized.
1. Time of change + bottle number
Each bottle is allocated a specific filling time.
External BC sync input
The system changes to the next bottle when
an external signal is received. The external
signal first has to be configured via the
binary input. The binary input can then be
selected as the source.
Assignment bin. output Selection
• No binary output
configured
• Binary output S:x
Factory setting
No binary output
configured
Assignment of the binary output to the
program cycle.
Use "SAVE" to save the subprogram setup. Then press "ESC" to return to the main program. A prompt to save the
program appears if you have not yet saved the subprogram. You can avoid saving the program by pressing "ESC".
Inputs Settings for the inputs can be made as
described in the "Inputs" section.
Bottle assignment(only possible
with multiple bottles)
This menu item appears when
more than one bottle is available,
regardless of the number of
subprograms.
Selection
• No bottle assignment
• Dynamical bottle
assignment
• Statical bottle
assignment
Factory setting
Dynamical bottle
assignment
No bottle assignment:
Each subprogram fills the same bottle until
the bottle is full. All the subprograms then
change to the next bottle. Only visible if
there is more than one subprogram.
Dynamical bottle assignment:
When the subprogram changes, the system
switches to the next empty bottle
Statical bottle assignment:
A table can be used to assign a subprogram
to each bottle
Via the "Bottle change" menu item, the bottle change can be configured after a certain time or number of
samples if bottle distribution with more than one bottle has been selected and either dynamic or static bottle
assignment has been selected.
Having selected Bottle assignmentStatical bottle assignment:
Bottle assignment table
Select a bottle and assign it a subprogram.