Operation Liquistation CSF33
82 Endress+Hauser
Menu/Setup/Sampling programs
Function Options Info
Repetition mode Selection
• Daily interval
• Weekly interval
• Days of week
Factory setting
Daily interval
Daily interval
Specify whether the subprogram should be
repeated every day.
Weekly interval
Specify whether the subprogram should be
repeated every week.
Days of week
Specify whether the subprogram should be
repeated on certain days of the week. -->
Select the days of the week in the
subsequent menu item.
Repetition interval
(only with Daily interval
and Weekly interval)
1 ... 999
Factory setting
1
Specify for how many days or weeks the
subprogram should be active.
Example: Repetition mode = daily interval
Repetition interval = 2
The subprogram is enabled every second day
from the start condition.
Having selected Enable subprogram Interval
Ensure activation Selection
• No
• Daily
• Weekly
Factory setting
No
Ensures that the subprogram is activated at
the specified intervals. If necessary, the
inactivity time is shortened by one day or
one week.
Start condition Selection
• No delay
• Date/Time
• Time
• No delay (sync)
Factory setting
No delay (sync)
The subprogram is started when the
program is enabled.
Date/Time
Set the start date and start time for enabling
the subprogram.
Time
Set the start time for enabling the
subprogram.
No delay (sync)
Only possible on program start Immediate
and with bottle assignment "Dynamic or
Static".
Start date 01.01.2000 ... 31.12.2099
Factory setting
DD.MM.YYYY
Set the start date for the 1st interval. The
format depends on the option configured
under general settings.
Start time 00:00:00 ... 23:59:59
Factory setting
00‐00:01 DD‐HH:MM
Set the time for the 1st interval. The format
depends on the option configured under
general settings.
Activity time 00-00:01 ... 31-00:00
DD-HH:MM
Factory setting
00‐00:01 DD‐HH:MM
Specify how long the subprogram should be
active in days, hours and minutes. The
subprogram always begins with an
activation.
Inactivity time 00-00:01 ... 31-00:00
DD-HH:MM
Factory setting
00‐00:01 DD‐HH:MM
Specify how long the subprogram should be
inactive in days, hours and minutes.
Sample at enable Selection
• Yes
• No
Factory setting
Yes
Specify whether the first sample should be
taken directly when the subprogram is
enabled. For example, with intervals, a
sample is taken at the start of every
activation interval.