Liquistation CSF33 Operation
Endress+Hauser 93
A
B
1
1
2
8.0
A0014126
 30 Event-controlled sampling
A pH value: 8.0
B pH value
1 Sampling into individual bottles when pH value > 8.0
2 Sampling into composite container when pH value < 8.0
Menu/Setup/Sampling programs/Setup program/New
Function Entry for sample program
Advanced
Program name TDTE 12+20
Bottle configuration 12x + 1x
Bottle volume left 1000 ml
Bottle volume right 2000 ml
Start condition Immediate
Stop condition Continuous
Bottle assignment Statical bottle assignment
Setup subprogram: New
Programpart TD1
Sampling mode Time paced CTCV
Sampling interval 15 min
Sampling volume 100 ml
Multiple bottles 1
Enable subprogram Immediate
Sample at enable Yes
Synchronize samplings To subprogram start
> "SAVE" > "ESC"
Setup subprogram: New
Programpart EE1
Sampling mode Time paced CTCV
Sampling interval 6 min
Sampling volume 100 ml
Multiple bottles Only having
selected Bottle assignment =
Statical bottle assignment in
the main program
Number of samples
Samples per bottle 10