Liquistation CSF33 Operation
Endress+Hauser 99
Menu/Setup/Inputs
Function Options Info
Input mode Selection
• Flow rate
• Parameter
• Current
Factory setting
Current
Select the input variable.
Flow rate
The input can be used as the source for
time/flow-paced or flow-paced sampling
programs.
Parameter
The input can be used as the source for limit
switches, logbooks and enabling and
disabling events for sampling programs.
Current
The input can be used as the source for limit
switches, logbooks and enabling and
disabling events for sampling programs. A
unit name cannot be specified.
Having selected Input mode Flow rate
Unit of flow Selection
• l/s
• m³/s
• m³/h
• m³/d
• cfs
• gpm
• gph
• mgd
Factory setting
l/s
Select the unit.
Unit of totalized flow Selection
• l
• m³
• cf
• gal
Factory setting
m³
Select the unit for the totalized flow.
Meas. value format Factory setting
#.#
Specify the number of decimal places for the
flow.
Minimum flow 0 to 10000 l/s
Factory setting
0 l/s
The set limit value prevents sampling if the
flow falls below this value (only for time/
flow-paced sampling).
Lower range value 0 to 10000 l/s
Factory setting
0 l/s
Enter a value for the start of the measuring
range. 0/4 mA is assigned to this value as
per your specifications.
Upper range value 0 to 10000 l/s
Factory setting
100000 l/s
Enter a value for the end of the measuring
range. 20 mA is assigned to this value as per
your specifications.
Damping 0 ... 60 s
Factory setting
0 s
The damping causes a floating average curve
of the measured values over the time
specified.
Totalized flow
The totalized flow is calculated when the program is started if you use a sampling program with
volume, flow-paced sampling or time/flow-paced sampling as the start condition. The samples are
taken based on this value. The current totalizer is used for calculating purposes if the totalized flow is
used as the measured value for an enabling or disabling event.
Current totalized flow - - - The totalized flow values are displayed.