Commissioning Liquisys M CPM223/253
72 Endress+Hauser
6.6.9 Interfaces
6.7 Communication
For devices with a communication interface, please also refer to the separate Operating
Instructions BA00208C/07/EN (HART) or BA00209C/07/EN (PROFIBUS).
Coding Field Setting range
(Factory settings, bold)
Display Info
I
INTERFACE
function group
Communication settings (only for device
version HART or PROFIBUS).
I1 Enter bus address
Address
HART: 0 to 15
or
PROFIBUS: 0 to 126
Each address may only be used once in a
network.
If a device address 0 is selected, the
current output is automatically set to 4
mA and the device is set to multi-drop
operation.
I2
Display of measuring
point
I
INTERFACE
I1
Address
126126
I2
@@@@@@@@
Tag