Micropilot FMR56, FMR57
Endress+Hauser 29
FOUNDATION Fieldbus
Device basic current 15 mA
Failure current FDE (Fault
Disconnection Electronic)
0 mA
FISCO
U
i
17.5 V
I
i
550 mA
P
i
5.5 W
C
i
5 nF
L
i
10 μH
Power supply failure
Configuration is retained in the HistoROM (EEPROM).
Error messages (incl. value of operated hours counter) are stored.
Potential equalization
No special measures for potential equalization are required.
If the device is designed for hazardous areas, observe the information in the documentation
"Safety Instructions" (XA).
Terminals
Without integrated overvoltage protection
Plug-in spring terminals for wire cross-sections 0.5 to 2.5 mm
2
(20 to 14 AWG)
With integrated overvoltage protection
Screw terminals for wire cross-sections 0.2 to 2.5 mm
2
(24 to 14 AWG)
Cable entries Connection of the power supply and signal cables
To be selected in feature 050 "Electrical connection":
Coupling M20, material depends on approval:
For non-Ex, ATEX, IECEx, NEPSI Ex ia/ic:
Plastic M20x1.5 for cable ⌀5 to 10 mm (0.2 to 0.39 in)
For Dust-Ex, FM IS, CSA IS, CSA GP, Ex ec:
For Ex db:
No cable gland available
Thread
½" NPT
G ½"
M20 × 1.5
M12 plug / 7/8" plug
Only available for non-Ex, Ex ic, Ex ia
Connection of remote display FHX50
Feature 030 "Display, operation" Cable entry for connection of FHX50
L: "Prepared for display FHX50 + M12 connection" M12 socket
M: "Prepared for display FHX50 + M16 cable gland, custom connection" M12 cable gland
N: "Prepared for display FHX50 + NPT1/2 thread, custom connection" Thread NPT1/2
Cable specification
Devices without integrated overvoltage protection
Pluggable spring-force terminals for wire cross-sections 0.5 to 2.5 mm
2
(20 to 14 AWG)
Devices with integrated overvoltage protection
Screw terminals for wire cross-sections 0.2 to 2.5 mm
2
(24 to 14 AWG)
For ambient temperature T
U
≥60 °C (140 °F): use cable for temperature T
U
+20 K.