Micropilot FMR60B PROFINET with Ethernet-APL Mounting
Endress+Hauser 13
D H
max
100 to 150 mm (4 to 6 in) 550 mm (22 in)
≥ 150 mm (6 in) 850 mm (34 in)
In the case of longer nozzles, reduced measuring performance must be expected.
Please note the following:
The end of the nozzle must be smooth and free from burrs.
The edge of the nozzle should be rounded.
Mapping must be performed.
Please contact the manufacturer's support department for applications with nozzles
that are higher than indicated in the table.
10.3.4 Integrated antenna, PEEK 40 mm (1.5 in)
Information concerning threaded connections
When screwing in, turn by the hex bolt only.
Tool: open-ended wrench 55 mm
Maximum permissible torque: 50 Nm (36 lbf ft)
Information about the mounting nozzle
The maximum nozzle length H
max
depends on the nozzle diameter D.
Maximum nozzle length H
max
as a function of the nozzle diameter D
ØD H
max
H
øD
max
40 to 50 mm (1.6 to 2 in) 400 mm (16 in)
50 to 80 mm (2 to 3.2 in) 550 mm (22 in)
80 to 100 mm (3.2 to 4 in) 850 mm (34 in)
100 to 150 mm (4 to 6 in) 1 050 mm (42 in)
≥150 mm (6 in) 1 600 mm (64 in)
In the case of longer nozzles, reduced measuring performance must be expected.
Please note the following:
The end of the nozzle must be smooth and free from burrs.
The edge of the nozzle should be rounded.
Mapping must be performed.
Please contact the manufacturer's support department for applications with nozzles
that are higher than indicated in the table.
10.3.5 Turning the housing
The housing can be rotated up to 380° by loosening the locking screw.