Micropilot FMR67
36 V. 1, Rev. 3, 11-04-2017 Endress+Hauser
Altitude according to
IEC61010-1 Ed.3
•
Generally up to 2 000 m (6 600 ft) above MSL.
• Above 2 000 m (6 600 ft) if the following conditions are met:
– Ordering feature 020 "Power supply; Output" = A, B, C, E or G (2-wire versions)
– Supply voltage U < 35 V
– Supply voltage of overvoltage category 1
Degree of protection
•
With closed housing tested according to:
– IP68, NEMA6P (24 h at 1.83 m under water surface)
– For plastic housing with transparent cover (display module): IP68 (24 h at 1.00 m under water
surface)
6)
– IP66, NEMA4X
• With open housing: IP20, NEMA1
• Display module: IP22, NEMA2
Degree of protection IP68 NEMA6P applies for M12 PROFIBUS PA plugs only when the
PROFIBUS cable is plugged in and is also rated
IP68 NEMA6P.
Vibration resistance
DIN EN 60068-2-64 IEC 60068-2-64 at
5 to 2 000 Hz: 1.5 (m/s
2
)
2
/Hz
A limited vibration resistance applies according
DIN EN 60068-2-64/IEC 60068-2-64 at
5 to 2 000 Hz of 0.39 (m/s
2
)
2
/Hz,
if the following order features are selected in the product structure at the same time:
• 040 ("Housing"): B ("GT18 two-chamber, 316L)
and
• 090 ("Seal"): A6 ("FKM Viton GLT, –40 to 200 °C (–40 to 392 °F), incl. gas-tight feedthrough")
Electromagnetic
compatibility (EMC)
Electromagnetic compatibility in accordance with all of the relevant requirements outlined in the
EN 61326 series
and NAMUR Recommendation EMC (NE 21). For details, please refer to the
Declaration of Conformity
7)
.
A normal device cable suffices if only the analog signal should be used. Use a shielded cable for
digital communication (HART/ PA/ FF).
Maximum measured error during EMC testing: <0.5 % of the span. By way of derogation, for devices
with a plastic housing and see-through lid (integrated display SD02 or SD03) the measured error can
be up to 2 % of the span in the event of strong electromagnetic radiation in the 1 to 2 GHz frequency
range.
6) This restriction is valid if the following options of the product structure have been selected at the same time: 030 ("Display, Operation") = C
("SD02") or E ("SD03"); 040 ("Housing") = A ("GT19").
7)
Available for download at www.de.endress.com.