Proline Fieldcheck Wiring
Endress+Hauser 15
3.4 Connecting to Promag 50, 53, 55 wall-mount housing
"
Caution!
Connect the ground terminals on the cables to the uninsulated housing components or to the
ground terminals in the connection compartment.
1. Remove the cover of the connection compartment (4).
2. Completely remove the terminal block (5, terminal nos. 4 to 8, 36+37) from the base.
"
Caution!
If the terminal block is not removed, a correct simulation/verification is not guaranteed.
3. Plug the Simubox cable connector into the amplifier socket (7).
4. Connect Fieldcheck to the current and frequency outputs of the transmitter
(terminal pairs 22/23, 24/25 or 26/27).
!
Note!
The configuration of the outputs depends on the product type and the order version of the
tested transmitter. To find out which terminal pairs are assigned to the outputs, see the
"Wiring" section of the flowmeter Operating Instructions.
5. Connect Fieldcheck to the service connector (6).
a0001699
Fig. 8: Connecting Fieldcheck to Promag 50, 53, 55 (wall-mount housing)
1 Fieldcheck flow simulator
2 Simubox
3 Transmitter wall-mount housing
4 Connection compartment cover
5 Terminal block
6 Service connector
7 Amplifier socket
1
2
22 2320 21 24
25
26
27
41
42
6
5
7
8
4
37 36
3
4
5
7
6
2
1