Wiring Proline Fieldcheck
24 Endress+Hauser
3.10 Connecting to Prowirl 72, 73
"
Caution!
Connect the ground terminals on the cables to the uninsulated housing components or to the
ground terminals in the connection compartment.
1. Remove the cover of the electronics compartment (3).
2. Remove the local display module (4) from the retainer rails (5).
3. Fit the local display module (4) with its left side onto the right retainer rail (5) (this secures the
local display module).
(Continued on next page)
a0001702
Fig. 17: Connecting Fieldcheck to Prowirl 72, 73
1 Fieldcheck flow simulator
2Simubox
3 Electronics compartment cover
4 Local display module
5 Local display module retainer rails
6 Fixing screw for cover of connection compartment
7 Connection compartment cover
8 Current output/power supply terminal
9 Pulse output terminal (optional)
10 Plastic cover
11 Amplifier socket
12 Service connector
2
1
5
7
6
4
3
7
10
11
12
8
12 34
++
--
9
10