Device functions - thermal measuring systems Proline Fieldcheck
74 Endress+Hauser
12.1.1 Group SIMULATION
Function group CONFIGURATION
Function description
FUNCTION SIMULATION CONFIGURATION
OPERATION MODE Use this function to select the operation mode for the simulation.
Users can freely configure a flow signal from 0.0 to the displayed maximum value.
Display:
CONTINUOUS
The flow behavior in the process has low dynamics.
a0001708
Q1 = Simulation value 1
FLOW SPECIFICATION Use this function to select which measurement is to be simulated.
Options:
MASS FLOW
CORRECTED VOLUME FLOW
Factory setting:
MASS FLOW
SIGNAL 1 LEVEL Use this function to enter the flow value which is to be simulated.
!
Note!
More information is provided in the OPERATION MODE function.
User input:
0.000 to 999999 (6-digit positive floating-point number)
!
Note!
The flow velocity must lie within the value range from 0.0 m/s to the displayed
maximum value.
Unit:
The unit configured in the measuring device is adopted.
Factory setting:
0.000