Proline Fieldcheck Wiring
Endress+Hauser 21
3.7 Connecting to t-mass 65 field housing
"
Caution!
Connect the ground terminals on the cables to the uninsulated housing components or to the
ground terminals in the connection compartment.
1. Remove the cover of the electronics compartment (4).
2. Pull down the local display (6).
3. Pull the sensor signal cable connector out of the amplifier socket (5).
4. Plug the Simubox cable connector into the amplifier socket (5).
5. Remove the cover of the connection compartment (7) from the transmitter housing.
6. Connect Fieldcheck to the current and frequency outputs of the transmitter
(terminal pairs 22/23, 24/25 or 26/27).
!
Note!
The configuration of the outputs depends on the product type and the order version of the
tested transmitter. To find out which terminal pairs are assigned to the outputs, see the
"Wiring" Section of the flowmeter Operating Instructions.
7. Connect Fieldcheck to the service connector (8).
A0013508
Fig. 14: Connecting Fieldcheck to t-mass 65 (field housing)
1 Fieldcheck flow simulator
2 Simubox
3 Aluminum transmitter
4 Electronics compartment cover
7 Connection compartment cover
5 Amplifier socket
6 Display
8 Service connector
E
-
+
3
5
6
2
1
27
25
23
21
2
1
26
24
22
20
L1 (L+)
N (L-)
8
7
4