Proline Fieldcheck Device functions - electromagnetic measuring systems
Endress+Hauser 51
FLOW SPECIFICATION Use this function to select which measurement is to be simulated.
Options:
VOLUME FLOW
MASS FLOW
!
Note!
"Calculated mass flow" simulation is only possible with Promag 53 type devices.
The value stored in the DENSITY VALUE function (DENSITY PARAMETER function
group of the device) is used for the calculation.
If a current input is available in the device and the density is assigned to it, mass flow
simulation is not possible.
Factory setting:
VOLUME FLOW
SIGNAL 1 LEVEL/
SIGNAL 2 LEVEL
Use this function to enter the flow value which is to be simulated.
Enter the second flow value which is to be simulated in SIGNAL 2 LEVEL if you have
selected the STEP or PROFILE operating mode.
!
Note!
More information is provided in the OPERATION MODE function.
User input:
0.000 to 999999 (6-digit floating-point number with sign)
!
Note!
The flow velocity must lie within the value range -12.5 m/s to +12.5 m/s.
Unit:
The unit configured in the measuring device is adopted.
Factory setting:
0.000
SIGNAL 1 PERIOD/
SIGNAL 2 PERIOD
Use this function to enter a duration for simulation signals 1 and 2 if you have selected
the STEP or PROFILE operating mode.
!
Note!
More information is provided in the OPERATION MODE function.
User input:
1 to 999 s (seconds)
Factory setting:
10 s
SIGNAL 1 PROFILE/
SIGNAL 2 PROFILE
For SIGNAL 1 PROFILE, enter the period of time for the change from simulation signal
1 to simulation signal 2. For SIGNAL 2 PROFILE, enter the period of time for the change
from simulation signal 2 to simulation signal 1.
!
Note!
This function is only available if the PROFILE setting was selected in the OPERATION
MODE function.
More information is provided in the OPERATION MODE function.
User input:
1 to 99 s (seconds)
Factory setting:
5 s
Function description
FUNCTION SIMULATION CONFIGURATION