Device functions - electromagnetic measuring systems Proline Fieldcheck
54 Endress+Hauser
Analyzing the results
During verification, the current verification test section is displayed in the 'HOME position'. The
results can be shown on the display once verification is complete. They can be filed in Fieldcheck
under a storage location number of your choice (up to max. 20) together with the measuring device
serial no. and the tag name. Via a connection with the FieldCare software, a report of the results
can later be printed out and the data can be archived.
Function group APPLICATION/OUTPUTS
During verification, the function and accuracy of a current output and a frequency output can be
checked. Fieldcheck determines the current and frequency outputs available in the measuring
device.
Function description
FUNCTION VERIFICATION APPLICATION/OUTPUTS
OUTPUTS Use this function to display the outputs determined in the flowmeter on the local display
with the related terminals and to select the outputs to be tested. The outputs selected
are displayed in white with a black background.
1. Using the
P
keys, select the outputs to be tested during the verification process.
2. Confirm the option selected with the
F
key.
!
Note!
It is not possible to select two outputs of the same type.
Start the verification process in the VERIFICATE function ä 57.
Two outputs of the same type are tested in two separate runs. After the first run, the
message "MORE OUTPUTS TO BE TESTED?" appears on the display.
If you want to test other outputs, Fieldcheck returns to the output selection area.
!
Note!
Having selected an output, make sure that the related terminal pair of the flowmeter is
connected to the Fieldcheck input before the verification process is started.
The following combinations of current and frequency outputs are possible for Proline
Promag:
•I
•I+I
•I+F
I+I+F
I+F+F
•I+P (only Promag 10)
!
Note!
Fieldbus interfaces cannot be tested.