Proline Fieldcheck Device functions - vortex measuring devices
Endress+Hauser 87
Function group LIMIT VALUES
You can specify the limit values for measured variables in this function group. If the deviation of a
verification step (see OPERATION, P. 38) lies within the configured limit values the result is
PASSED; if it lies outside of this range the result is FAILED. For quick and unambiguous
identification, the FAILED results are shown in white with a black background on the display.
Function description
FUNCTION VERIFICATION LIMIT VALUES
DEVIATION FLOW
(for liquid
measurement)
User input:
0.5 to 10.00 %
Factory setting:
0.5 %
DEVIATION FLOW
(when measuring gas or
steam)
User input:
0.5 to 10.00 %
Factory setting:
0.5 %
DEVIATION CURRENT
OUTPUT
This function is only available if the current output was selected in the OUTPUTS
function.
User input:
0.02 to 1.00 mA
Factory setting:
0.05 mA
DEVIATION
TEMPERATURE
(only Prowirl 73)
User input:
2.00 to 10.00 (for the units °C and K)
4.00 to 10.00 (for the units °F and R)
Unit:
The unit configured in the measuring device is adopted.
Factory setting:
4.0 (for the units °C and K)
8.0 (for the units °F and R)