Proline Fieldcheck Device functions - vortex measuring devices
Endress+Hauser 89
Function group OPERATION
Use this function group to specify the scope of the verification and to start/stop the verification.
Function description
FUNCTION VERIFICATION OPERATION
VERIFICATION Use this function to define the scope of the test.
Options:
TRANSM. + SENSOR
TRANSMITTER
SENSOR
Factory setting:
TRANSM. + SENSOR
VERIFICATE Use this function to control the verification. After completion of the verification, the
question STORE DATA? appears in this function.
Options:
START
The verification is started and the message "VERIFICATION ACTIVE" appears on the
display. At the end of the verification, the message "RESULT: PASSED/FAILED - STORE
DATA?" appears on the display.
STOP
Via the STOP option the active verification can be aborted.
STORE DATA?
The results of the verification can be saved using the STORE DATA? option.
Factory setting:
START
STORAGE LOCATION Use this function to define where the results of the verification are to be stored.
User input:
1 to 20
Factory setting:
First available storage location
!
Note!
If there is no storage location available or the selected storage location is occupied, the
warning "OVERWRITE STORAGE LOCATION? Y/N" is output before saving.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser Proline Fieldcheck and is the answer not in the manual?

Endress+Hauser Proline Fieldcheck Specifications

General IconGeneral
BrandEndress+Hauser
ModelProline Fieldcheck
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals