Proline Fieldcheck Device functions - ultrasonic measuring systems
Endress+Hauser 97
Function group OPERATION
Use this function group to verify the measuring point.
Function description
FUNCTION VERIFICATION OPERATION
VERIFICATION Use this function to define the scope of the verification.
!
Note!
• Verification of the current and frequency outputs of the transmitter can only be
achieved with the TRANSM. + SENSOR or TRANSMITTER option.
• If TRANSM. + SENSOR or SENSOR is selected, the sensor test is carried out at the end
of the verification ä 93.
Options:
TRANSM. + SENSOR (only "clamp-on")
TRANSMITTER
SENSOR (only "clamp-on")
Factory setting:
TRANSM. + SENSOR (only "clamp-on")
VERIFICATE Use this function to control the verification. After completion of the verification, the
question STORE DATA? appears in this function.
Options:
START
The verification is started and the message "VERIFICATION ACTIVE" appears on the
display. At the end of the verification, the message "RESULT: PASSED/FAILED - STORE
DATA?" appears on the display.
STOP
Via this option the active verification can be aborted.
STORE DATA?
The results of the verification can be saved using this option.
Factory setting:
START
STORAGE LOCATION Use this function to define where the results of the verification are to be stored.
User input:
1 to 20
Factory setting:
First available storage location
!
Note!
If there is no storage location available or the selected storage location is occupied, the
warning OVERWRITE STORAGE LOCATION? Y/N is output before saving.