EasyManua.ls Logo

Endress+Hauser Proline Prosonic Flow B 200 - Page 91

Endress+Hauser Proline Prosonic Flow B 200
156 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Proline Prosonic Flow B 200 HART Commissioning
Endress+Hauser 91
Parameter overview with brief description
Parameter Prerequsite Description User interface / 
Selection
Factory setting
Operating time Indicates how long the device
has been in operation.
Days (d), hours (h),
minutes (m) and
seconds (s)
Last backup A local display is provided. Indicates when the last data
backup was saved to the
display module.
Days (d), hours (h),
minutes (m) and
seconds (s)
Configuration management A local display is provided. Select action for managing the
device data in the display
module.
Cancel
Execute backup
Restore
Duplicate
Compare
Clear backup data
Cancel
Comparison result A local display is provided. Comparison between present
device data and display
backup.
Settings identical
Settings not
identical
No backup
available
Backup settings
corrupt
Check not done
Dataset
incompatible
Check not done
10.6.1 Function scope of the "Configuration management" parameter
Options Description
Execute backup The current device configuration is backed up from the integrated HistoROM to the
device's display module. The backup copy includes the transmitter data of the
device.
Restore The last backup copy of the device configuration is restored from the display
module to the device's integrated HistoROM. The backup copy includes the
transmitter data of the device.
Duplicate The transmitter configuration from another device is duplicated to the device using
the display module.
Compare The device configuration saved in the display module is compared with the current
device configuration of the integrated HistoROM.
Clear backup data The backup copy of the device configuration is deleted from the display module of
the device.
Integrated HistoROM
A HistoROM is a "non-volatile" device memory in the form of an EEPROM.
While this action is in progress, the configuration cannot be edited via the local
display and a message on the processing status appears on the display.
10.7 Simulation
The Simulation submenu enables you to simulate, without a real flow situation, various
process variables in the process and the device alarm mode and to verify downstream
signal chains (switching valves or closed-control loops).

Other manuals for Endress+Hauser Proline Prosonic Flow B 200

Related product manuals