Proline Prosonic Flow P 500
40 Endress+Hauser
4. Screw the nuts onto the U-shaped screws.
5. Position the sensor holder correctly and tighten the nuts uniformly.
1
A0043369
 27 Holder with U-shaped screws
1 Sensor holder
L
CAUTION
Risk of damaging plastic or glass pipes if the nuts on the U-shaped screws are tightened too
much!
The use of a metal half-shell (on the opposite side of the sensor) is recommended for plastic or
glass pipes.
The visible measuring pipe surface must be clean (free from flaking paint and/or rust) to ensure
good acoustic contact.
Sensor holder with strapping bands (small nominal diameters)
Can be used for
Measuring devices with measuring range DN 15 to 65 (½ to 2½")
Mounting on pipes DN > 32 (1¼")
Procedure:
1. Disconnect the sensor from the sensor holder.
2. Position the sensor holder on the measuring pipe.