Proline Prosonic Flow P 500
Endress+Hauser 41
3. Wrap the strapping bands around the sensor holder and measuring pipe without twisting
them.
1
A0043371
 28 Positioning the sensor holder and mounting the strapping bands
1 Sensor holder
4. Guide the strapping bands through the strapping band locks.
5. Tighten the strapping bands as tightly as possible by hand.
6. Set the sensor holder to the desired position.
7. Push down the tensioning screw and tighten the strapping bands so they cannot slip.
1
A0043372
 29 Tightening the tensioning screws of the strapping bands
1 Sensor holder
8. If necessary, shorten the strapping bands and trim the cut edges.
L
WARNING
Risk of injury!
To avoid sharp edges, trim the cut edges after shortening the strapping bands. Wear suitable
gloves and protective goggles.
The visible measuring pipe surface must be clean (free from flaking paint and/or rust) to ensure
good acoustic contact.