Proline Prosonic Flow P 500
Endress+Hauser 49
Installation for measurement via 2 traverses
Requirements
The installation distance is known →  39
Strapping bands are pre-assembled
Material
The following material is required for mounting:
Two strapping bands incl. mounting bolts and centering plates where necessary (already pre-
assembled →  42, →  43)
A mounting rail to position the strapping bands:
Short rail up to DN 200 (8")
Long rail up to DN 600 (24")
No rail > DN 600 (24"), as distance measured by sensor distance between the mounting bolts
Two mounting rail holders
Two sensor holders
Coupling medium (coupling pad or coupling gel) for an acoustic connection between the sensor
and pipe
Two sensors incl. connecting cables
Open-ended wrench (13 mm)
Screwdriver
Procedure:
1. Position the strapping bands using the mounting rail [only DN50 to 600 (2 to 24"), for larger
nominal diameters, measure the distance between the center of the strap bolts directly]: Fit
the mounting rail with the bore identified by the letter (from the Result sensor distance /
measuring aid parameter) over the mounting bolt of strapping band 1 that is fixed in place.
Position the adjustable strapping band 2 and fit the mounting rail with the bore identified by
the numerical value over the mounting bolt.
G
22
A0043384
 42 Determining the distance according to the mounting rail (e.g. G22)
2. Tighten strapping band 2 so that it cannot slip.
3. Remove the mounting rail from the mounting bolt.
4. Fit the sensor holders over the individual mounting bolts and tighten securely with the locking
nut.
5. Screw the mounting rail holders onto the sensor holder.