Proline Prosonic Flow P 500
Endress+Hauser 9
Pipe wall thickness [mm (in)]
Kinematic viscosity cSt [mm
2
/s]
0 < v ≤ 10 10 < v ≤ 100 100 < v ≤ 1000
Transducer frequency (sensor version / number of traverses)
1)
5.9 to 10.0 (0.23 to 0.39) 2 MHz (C-200 / 2) 1 MHz (C-100 / 2) 0.5 MHz (C-050 / 2)
>10.0 (0.39) 1 MHz (C-100 / 2) 1 MHz (C-100 / 1) 0.5 MHz (C-050 / 1)
1) The table shows a typical selection. In critical situations, the optimum sensor type may differ from these recommendations.
Measuring pipe material: plastic
Nominal diameter [mm (")]
Kinematic viscosity cSt [mm
2
/s]
0 < v ≤ 10 10 < v ≤ 100 100 < v ≤ 1000
Transducer frequency (sensor version / number of traverses)
1)
15 to 50 (¹⁄₂ to 2) 5 MHz (C-500 / 2) 5 MHz (C-500 / 2) 5 MHz (C-500 / 2)
50 to 80 (2 to 3) 2 MHz (C-200 / 2) 1 MHz (C-100 / 2) 0.5 MHz (C-050 / 2)
80 to 150 (3 to 6) 1 MHz (C-100 / 2) 1 MHz (C-100 / 2) 0.5 MHz (C-050 / 2)
150 to 200 (6 to 8) 1 MHz (C-100 / 2) 0.5 MHz (C-050 / 2) 0.5 MHz (C-050 / 2)
200 to 300 (8 to 12) 1 MHz (C-100 / 2) 0.5 MHz (C-050 / 2) 0.5 MHz (C-050 / 2)
300 to 400 (12 to 16) 1 MHz (C-100 / 1) 0.5 MHz (C-050 / 2) 0.5 MHz (C-050 / 1)
400 to 500 (16 to 20) 1 MHz (C-100 / 1) 0.5 MHz (C-050 / 1) 0.5 MHz (C-050 / 1)
500 to 1 000 (20 to 40) 0.5 MHz (C-050 / 1) 0.5 MHz (C-050 / 1) –
1 000 to 4 000 (40 to 160) 0.3 MHz (C-030 / 1) – –
1) The table shows a typical selection. In critical situations, the optimum sensor type may differ from these recommendations.
Measuring pipe material: glass-fiber reinforced plastic
Nominal diameter [mm (")]
Kinematic viscosity cSt [mm
2
/s]
0 < v ≤ 10 10 < v ≤ 100 100 < v ≤ 1000
Transducer frequency (sensor version / number of traverses)
1)
15 to 50 (¹⁄₂ to 2) 5 MHz (C-500 / 2) 5 MHz (C-500 / 2) 5 MHz (C-500 / 2)
50 to 80 (2 to 3) 1 MHz (C-100 / 2) 0.5 MHz (C-050 / 2) 0.5 MHz (C-050 / 1)
80 to 150 (3 to 6) 1 MHz (C-100 / 2) 0.5 MHz (C-050 / 1) 0.5 MHz (C-050 / 1)
150 to 200 (6 to 8) 0.5 MHz (C-050 / 2) 0.5 MHz (C-050 / 1) –
200 to 300 (8 to 12) 0.5 MHz (C-050 / 2) 0.5 MHz (C-050 / 1) –
300 to 400 (12 to 16) 0.5 MHz (C-050 / 2) 0.5 MHz (C-050 / 1) –
400 to 500 (16 to 20) 0.5 MHz (C-050 / 1) – –
500 to 1 000 (20 to 40) 0.5 MHz (C-050 / 1) – –
1 000 to 4 000 (40 to 160) 0.3 MHz (C-030 / 1) – –
1) The table shows a typical selection. In critical situations, the optimum sensor type may differ from these recommendations.
• If clamp-on sensors are used, a 2 traverse-type installation is recommended. This is the
easiest and most convenient type of installation, particularly for measuring devices whose
pipe can only be accessed from one side.
• A 1 traverse installation is recommended for the following installation conditions:
• Certain plastic pipes with a wall thickness >4 mm (0.16 in)
• Pipes made of composite materials (e.g. glass-fiber reinforced plastic)
• Lined pipes
• Applications with media with high acoustic damping