Prosonic Flow 92 3 Installation
Endress+Hauser 19
For sensors P - DN 50...300
1. Push one of the supplied threaded bolts on the tensioning band (or both bolts in the
case of sound velocity measurement).
2. Run the tensioning band around the pipe without twisting it and push the end
through the tensioning band lock (make sure that the screw is pushed up).
3. By hand, make the tensioning band as tight as possible.
4. Push the screw down and tighten the tensioning band with a screwdriver so that it
cannot slip.
5. If so desired, shorten the tensioning band to the desired length.
" Caution!
Risk of injury! When shortening the tensioning band, avoid sharp edges.
Fig. 11: Tensioning band installation for DN 50...300
F06-9xxxxxxx-17-05-06-xx-000