Prothermo NMT81 Operation
Endress+Hauser 85
Present calibration
Present calibration shows the result of the current performed calibration as follows.
Confirm that the all values are shown appropriately.
A0044917
 89 Present calibration screen
10.2.4 HART output
Configuration
Configuration has four items to be set or confirmed as follows.
A0044915
 90 Configuration screen
HART address: Shows the communication address of the device and NMT81 is usually
set to 2.
Setting range: 0 to 63
HART short tag: Defines the name of the device to identify the device easily in the
master device.
Allowed characters type: 8 characters with A to Z, 0 to 9, and certain special characters
(***).
Device tag (long tag): Defines the name of the device to identify the device easily.
Allowed characters type: 0 to 32 characters with A to Z, 0 to 9, and certain special
characters (***).
No. of preambles: Defines the number of the bit when communicating with HART.
NMT81 is usually set to 5.
Setting range: 5 to 20
If you select "with tag" when ordering the device, you can apply names to both short
or long tags.
When changing the No. of preambles, the HART master must be set to the same
number as the NMT81.