Installation RIA45
10 Endress+Hauser
4.4 Installation procedure
The panel cutout required measures 92 mm (3.62 in) x 45 mm (1.77 in)
A0010404
 2 Installation in the panel
1. Screw the threaded rods (pos. 1) into the positions provided on the mounting frame
(pos. 2). Four opposing screw positions (pos. 3/4) are available for this purpose.
2. Push the device through the panel cutout from the front.
3. To secure the casing in the panel, hold the device in a horizontal position and push
the frame (pos. 2), with the threaded rods screwed in, over the casing until the frame
locks into position (1.).
4. Then tighten the threaded rods to fix the device in place (2.).
5. or the Ex option, install the distance piece (pos. 6) for the input terminals.
To disassemble the device, the mounting frame can be unlocked at the locking elements
(pos. 5) and then removed.
4.5 Post-installation check
Is the sealing which is spray-applied to the casing undamaged?
Is the mounting frame securely engaged on the housing of the device?
Are the threaded rods tightened?
Is the device positioned in the center of the panel cutout?
Is the distance piece installed (Ex option)?