RIA45 Commissioning
Endress+Hauser 37
List of error codes, see Troubleshooting →  40.
Under certain conditions events saved over the past 30 minutes might be lost.
7.5.5 Operating hours counter
The device has an internal operating hours counter which also acts as the reference for
diagnostic events.
The operating hours can be found in the ‘Diagnostics’ → ‘Operating time’ menu item. This
information cannot be reset or changed.
7.5.6 Device reset
The following reset categories are available for a device reset.
‘Expert’ → ‘System’ → ‘Reset’ → ‘Factory reset’: reset all the parameters to the as-
delivered state; all the configured parameters are overwritten.
Any user code already defined is overwritten!!! When operation is locked by a user
code, this is indicated by a lock symbol on the display.
‘Expert’ → ‘System’ → ‘Reset’ → ‘User reset’: parameters are loaded and configured in
accordance with the user setup saved; configuration currently configured or factory
settings are overwritten by the user setup.
Any user code already defined is overwritten by the user code defined in the user
setup!!! If no user code was saved in the user setup, the device is no longer locked.
When operation is locked by a user code, this is indicated by a lock symbol on the
display.