Soliwave FQR56/FDR56
Installation
Endress+Hauser 19
DimensionsL1toL3,dependingonthehousingandelectricalconnection:
Housing
Electrical connection
M20 gland ½ NPT thread
Connector
Binder M12
Connector
Harting HAN8D
F16, dimension L1 114 (4.49) 111 (4.37) 101 (3.98) 133 (5.24)
F15, dimension L2 93 (3.66) 94 (3.70) 88 (3.46) 120 (4.72)
F34, dimension L3 116 (4.57) 117 (4.60) 111 (4.37) 143 (5.63)
5�2 Installing the device
BoththeFQR56andtheFDR56haveastandardthread(R1½inaccordancewithEN10226,
1½ NPT in accordance with ANSI/ASME B1.20.1 or G 1½ in accordance with ISO 228-1) as
theprocessconnection,makingthemeasytoinstallinexistingcontainersleevesornozzles.
5�2�1 Installation with a self-sealing connecting thread (R 1½ and 1½ NPT)
10 Installationwithaself-sealingconnectingthread