Soliwave FQR56/FDR56
Commissioning
36
Endress+Hauser
8�3�2 Calibration with maximum covered path
This automatic setup function can be used if the path is maximum covered.
Move the encoding switch to position 2
→Displayofthefunctionnumber
Aer2seconds:Displayofthecurrentsignalstrength,example:
Simultaneously press the and keys on the device with maximum covered path
→Automaticcalibrationiscarriedout
→Displayofthesignalstrengthwithmaximumcoveredpath
Move the encoding switch to the initial position 0
→Displayofthecurrentsignalstrength
Optional: Additional manual calibration with free or minimum covered path (function 3)
Formostapplicationstheautomaticsetupissucient.Themanualsetupcanbeusedtoadapt
the microwave barrier individually to the application or medium.
IfaeranautomaticsetupwithCoveredpath(function2)thecondition"freepath"cannotbe
detected safely (switching point of the limit detection is not exceeded with free or minimum
coveredpath),thesensitivitymustbeincreasedwiththismanualsetupfunction3.
Move the encoding switch to position 3
→Displayofthefunctionnumber
Aer2seconds:Displayofthecurrentsignalstrength,example:
Press the
or key on the device to achieve an increase or reduction of the signal
strength display with free or minimum covered
→Displayofthesignalstrengthwithfreeorminimumcoveredpath(all10LEDs
illuminate)
Move the encoding switch to the initial position 0
→Displayofthecurrentsignalstrength
•TheSoliwavemicrowavebarrieriscalibrated,ifwithfreeorminimumcoveredpath
a) and devices with switching output the switching point (LED 5) is exceeded safely or
b) on devices with current output a minimum of 6 LEDs are lit (ideally all 10 LEDs are lit).
A repeated automatic calibration (function 1 or function 2) resets any calibration carried out
earlier.