EasyManuals Logo

Endress+Hauser SpectraSensors 2-Pack Analyzer System User Manual

Endress+Hauser SpectraSensors 2-Pack Analyzer System
121 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #47 background imageLoading...
Page #47 background image
Installation
Hardware Installation and Maintenance Manual 3
17
Tochange4‐20mAboardfromsourcetosink(AnalyzerAorB)
1. Disconnect power from the analyzer and open the electronics
enclosure cover. Take care not to disturb the electrical assembly
inside.
2. Locate the relay control board in the upper right of the electronics
enclosure, as shown in Figure 1–6 on page 1–11 or Figure 1–7 on
page 1–12 (Analyzer A) or Figure 1–8 on page 1–13 or Figure 1–9
on page 1–14 (Analyzer B).
3. Remove the jumper (JMP1), shown in Figure 3–8 on page 3–17,
connecting the center pin to point “A.
4. For 4-20 mA sink, carefully replace the jumper to connect the center
hole with point “P.
5. Reconnect power to the analyzer. Confirm the 4 mA (minimum) and
20 mA (maximum) points.
6. Close and tighten the analyzer electronics enclosure cover.
Needle nose pliers may be required to remove the jumper.
For Analyzer A, see “To calibrate the analog output (Analyzer
A)” on page 3-18. For the Analyzer B, refer to “To calibrate the
analog output (Analyzer B)” on page 3-18.
Figure 3–8 Analyzer 4-20 mA board
JMP1POTENTIOMETERS

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser SpectraSensors 2-Pack Analyzer System and is the answer not in the manual?

Endress+Hauser SpectraSensors 2-Pack Analyzer System Specifications

General IconGeneral
BrandEndress+Hauser
ModelSpectraSensors 2-Pack Analyzer System
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals