SS2100i-1 and SS2100i-2 Analyzer
44 4900002288 rev. B 1-13-21
The entire analyzer system is calibrated for operation at the
enclosure temperature and sample flow rate specified.
Measurements should be considered valid only when the enclosure
is at the specified temperature and sample flow rate. After opening
the sample system enclosure door to check settings, allow at least
1 to 2 hours for the temperature to re-stabilize before validating.
Figure 4–2 Typical sample system for differential
measurement with permeation tube validation
capability
PERMEATION
DRYER
TUBE
SAMPLE
SUPPLY
SAMPLE
RETURN
FLOW SHOWN IN THE
‘OFF’ POSITION
CELL 28 m