EasyManua.ls Logo

Endress+Hauser Waterpilot FMX21 - Inserting the Cable into the RIA15 Field Housing

Endress+Hauser Waterpilot FMX21
48 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Waterpilot FMX21 Installation
Endress+Hauser 21
5.8 Inserting the cable into the RIA15 field housing
TX20
2.
3.
4.
1.
A0017830
Inserting the cable, field housing, connection without transmitter power supply (example)
1. Release the housing screws
2. Open the housing
3. Open the cable gland (M16) and insert the cable
4. Connect the cable including the functional grounding and close the cable gland
Compensation of the atmospheric pressure must be ensured for the installation. A black,
vented cable gland is supplied for this purpose.
If using the communication resistor module in the RIA15, the cable of the FMX21 must
be inserted into the right gland when connecting the FMX21 so that the integrated
pressure compensation tube is not pinched.

Table of Contents