MEASERR is the last measured time offset of the CDMA Subsystem to CDMA while
locked, in seconds.
TIMEDEV is the time deviation (TDEV) of the offset measurements in seconds. The tau
associated with this measurement is one second, which is the update interval
of the position xes received from the CDMA Receiver.
AGERATE is the regression-computed system oscillator ageing rate per day (several-hour
delay before the rst measurements are displayed).
TAU is the system oscillator control loop averaging time constant, in seconds. It’s
value is automatically adjusted to maintain optimum system clock offset and
stability.
EFCDAC is the system oscillator Electronic Frequency Control 20-bit DAC value. The
system automatically sets this value to remove frequency errors. Values may
range from 0 to 1048575. Values close to the maximum or minimum will set the
DAC fault ag that will appear in the fault status display. The Time/Status dis
play will also indicate a fault condition.
TEMP is the chassis internal temperature in °C.
Below is an example of a typical response to this command:
Command: oscctrlstat
Sonoma reply:
20130117.00:23:10 LKD 0 6.26e-09 -6.26000e-09 1.25e-09 -6.93e-13 1955.3
524281 +50.750
passwd
This command is used to change the password for the user that you are logged in as. It affects the
serial port, SSH, Telnet and HTTPS. passwd is a Linux command that is also described in Appendix
C - Helpful Linux Information.
Command: passwd
Sonoma reply: Interactive script is started.
ptpconfig0 and ptpconfig1 (Optional)
These commands are only available if the Precision Time Protocol (PTP) option has been installed.
Refer to Chapter 4 - PTP/IEEE-1588 for more information.
ptpstat0 and ptpstat1 (Optional)
These commands are only available if the Precision Time Protocol (PTP) option has been installed.
Refer to Chapter 4 - PTP/IEEE-1588 for more information.