12 • Information on electromagnetic compatibility (EMC)
80114-770_V23802
2021-01
HF emissions CISPR 11 Class A The properties of this device in terms of its emissions
mean it can only be used in medical facilities that are
connected to supply systems specifically provided for
that purpose (usually supplied via isolating transform
-
ers). For domestic use (for which class B is usually
required as per CISPR
11), this device may not offer
adequate protection against radio services. The user
may need to take corrective measures such as relocat
-
ing or reorienting the device.
Harmonic emissions IEC 61000-3-2 Class A
Voltage fluctuations/flicker emissions
IEC
61000-3-3
Complies
Guidance and manufacturer's declaration - electromagnetic immunity
The equipment is intended for use in the electromagnetic environment specified below. The customer or the user of the
equipment should ensure that it is used in such an environment.
Immunity test IEC 60601-1-2 test level Compliance level Electromagnetic environment
- guidance
Discharge of static
electricity (ESD) in
accordance with
IEC 61000-4-2
±8 kV contact discharge
±15 kV air discharge
±8 kV contact discharge
±15 kV air discharge
The floor should be made from
wood or concrete or be covered
with ceramic tiles. If the floor is
covered with non-conductive
synthetic material, the relative
humidity must be at least 30%.
Electrical fast tran-
sient/burst IEC 61000-
4-4
±2 kV for power supply lines
±1 kV for input/output lines
±2 kV for power supply lines
±1 kV for input/output lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge IEC 61000-4-5 ±1 kV differential mode
±2 kV common mode
±1 kV differential mode
±2 kV common mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short-
term interruptions and
voltage fluctuations on
power supply input
lines as per
IEC 61000-4-11
0% U
T
for 0.5 cycle,
at 0, 45, 90, 135, 180, 225,
270 and 315 degrees
0% U
T
for 1 cycle,
single-phase at 0 degrees
70% U
T
for 25/30 cycles,
single-phase at 0 degrees
0% U
T
for 0.5 cycle,
at 0, 45, 90, 135, 180, 225,
270 and 315 degrees
0% U
T
for 1 cycle,
single-phase at 0 degrees
70% U
T
for 25/30 cycles,
single-phase at 0 degrees
Mains power quality should be
that of a typical commercial or
hospital environment.
If the user of the equipment
requires continued operation
during power mains interrup
-
tions, it is recommended that
the equipment be powered
from an uninterruptible power
supply or a battery.
Voltage monitoring as
per IEC 61000-4-11
0% U
T
for 250/300 cycles
(50/60
Hz)
0% U
T
for 250/300 cycles
(50/60
Hz)
Power frequency (50/
60 Hz) magnetic field
as per IEC 61000-4-8
30 A/m 30 A/m Power frequency magnetic
fields should be at levels char
-
acteristic of a typical location in
a typical commercial or hospital
environment.
Note: U
T
is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer's declaration - electromagnetic emissions