8 Installation
40 / 70
80114-770_V23802
2021-01
NOTICE
Insufficient acclimatization time, unsuitable temperature during
acclimatization
If the device was stored or transported below or above a certain tem-
perature, it will take a certain time and temperature to acclimatize.
If you do not observe the rules, the device can sustain damage and
fail.
Acclimatize the device according to the rules in the Technical Da-
ta.
NOTICE
Overheating of the device due to poor ventilation
If ventilation is poor, the device can overheat, sustain damage, and
fail.
Install the device in such a way that there is an unobstructed cir-
culation of air around the housing. Installation in confined wall re-
cesses is prohibited.
WARNING
Penetration of fluid into the device during operation or when
cleaning the device
Risk of electric shock to the medical personnel.
The unit may become damaged and fail.
Make sure no liquid can penetrate the device.
Do not place vessels containing liquids on top of the device.
If fluid has penetrated into the device, take the device out of op-
eration immediately. You can only start using the device again
once it has been checked by a service technician.
State of the unit and accessories
WARNING
Damaged device, damaged accessories, modified device, and
modified accessories
Risk of burns and injury to the patient and medical personnel! Risk of
damage to property.
Check the device and accessories for damage every time before
using them (e.g. footswitch, cords of instruments and the return
electrode, equipment cart).
You must not use damaged equipment or damaged accessories.
Replace defective accessories.
If the equipment or equipment cart is damaged, please contact
our customer service.
For your safety and that of the patient: Never attempt to perform
repairs or make modifications yourself. Any modification will in-
validate liability on the part of Erbe Elektromedizin GmbH.

Table of Contents

Other manuals for Erbe APC 3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Erbe APC 3 and is the answer not in the manual?

Erbe APC 3 Specifications

General IconGeneral
BrandErbe
ModelAPC 3
CategoryMedical Equipment
LanguageEnglish

Related product manuals