9 • Cleaning and disinfection
80114-770_V23802
2021-01
Chapter 9
Cleaning and disinfection
Safety Instructions
Connection of unit / equipment cart and power supply during
cleaning and disinfection
Risk of electric shock to the medical personnel!
Switch off the device. Unplug the power cord of the device/equip-
ment cart.
The unit is contaminated
Danger of infection to the patient and medical personnel.
Follow the instructions for cleaning and disinfecting the unit.
Alternate use of disinfectant solutions based on different active
ingredients
The agents may affect one another. The disinfection effect of the dis-
infectant solution may be weakened. Danger of infection to the pa-
tient and medical personnel.
The housing plastic may become brittle and break. A color reaction
may occur with plastics.
Do not use these substances alternately.
Flammable detergents and disinfectants, flammable solvents in
adhesives used on the patient and on the device / equipment cart
Risk of fire and explosion to the patient and medical personnel! Risk
of damage to property.
Use products that are not flammable.
If the use of flammable products is unavoidable, proceed as fol-
lows:
Allow the products to evaporate completely before switching on
the device.
Check whether flammable liquids have accumulated under the
patient, in body recesses such as the navel, or in body cavities
such as the vagina. Remove any liquids before performing elec-
trosurgery.