Ambient conditions
Ignition of anesthetics, skin cleansers, and disinfectants in poten-
tially explosive atmospheres
If you place the device in a potentially explosive atmosphere, anes-
thetics, skin cleansers, and disinfectants can ignite.
Risk of fire and explosion to the patient and medical personnel! Risk
of damage to property.
Do not place the device in potentially explosive atmospheres.
Interference with the unit from portable and mobile HF telecom-
munications equipment (e.g. mobile phones, WLAN equipment)
Electromagnetic waves emitted by portable and mobile HF telecom-
munications equipment may affect the unit.
Risk of injury to the patient.
The unit may fail or not perform properly.
When using portable and mobile HF telecommunications devices,
including their accessories, there must be a distance of at least
30 cm between them and the device and its cords.
Unsuitable temperature or level of humidity during operation
If you operate the equipment at an unsuitable temperature or level of
humidity, it may sustain damage, fail, or not perform properly.
Operate the equipment at a suitable temperature and level of hu-
midity. You will find the tolerances for temperature and humidity
in the Technical Data.
If other ambient conditions must be observed for operation of the
equipment, you will also find them in the Technical Data.
Unsuitable temperature or humidity in transit or storage
If you transport or store the equipment at an unsuitable temperature
or level of humidity, it may sustain damage and fail.
Transport and store the equipment at a suitable temperature and
level of humidity. You will find the tolerances for temperature and
humidity in the Technical Data.
If other ambient conditions must be observed for transport and
storage of the equipment, you will also find them in the Technical
Data.