- 19 -
correct cuff position (tab)
Caution
• Patient Hazard •
Apply the cu directly on the skin. Make sure that rolled up
sleeves do not impede blood circulation in the upper arm.
Loose cus will cause erroneous measurements; overtight cus
may constrict blood vessels or cause skin lesions and hemato-
mas.
• Incorrect Measurements •
A loose cu would degrade the accuracy of the measurement.
Therefore, the computer aborts the measurement, if a mini-
mum pressure is not attained within a few seconds.
Warning
• Patient Hazard •
If, by accident, an excessive pressure builds up inside the cu ,
either remove the cu immediately from the arm or disconnect
the cu tubing from the control terminal.
The same measures are recommended, if the cu does not
deate correctly.
When you close the Velcro strap, check that the metal
clasp (a) is inside the marked index range (b), and not
outside.
The cu tab must be located below the metal clasp (see
illustration at right).
CheCKinG the Cuff tuBinG
Check that the cu tubing does not knock against the
patient's knee, when the patient is pedalling and the hand
is on the handlebar.
Secure the cu tubing with the Velcro tape attached to the
handlebar.
Instruct your patient to move as little as possible during
a blood pressure measurement and, in particular, to avoid
excessive contractions of the muscles in the upper arm.
distance between Knee and tubing