- 41 -
↑
↓
↑
↓
settinGs for Control terminals K
Some of the device settings are congurable to meet spe‑
cic requirements. The settings will be saved and remain
stored even when the ergometer is switched o.
Select SETTINGS to display the conguration menu.
When all changes have been made, you can exit the con‑
guration menu with the key.
Use the softkeys (↑↓) to position the bar cursor on the pa‑
rameter to change and conrm the selection with SELECT.
Default moDe
In this menu you choose the default mode activated when
the ergometer is turned on. When rst turned on, the
ergometer will display this menu.
Use the softkeys (↑↓) to position the bar cursor on your
preferred default mode and save the selection with SELECT.
ProtoCols
Protocols 6 ‑ 15 are user‑programmable (protocols 1 ‑ 5
are xed, see Appendix for protocol parameter details).
Standard values for the following parameters can be
entered:
‑ initial load
‑ stage time
‑ load increment (load increase with each stage)
‑ NIBP lead time (blood pressure measurement)
‑ recovery load
‑ recovery time
Use the softkeys (↑↓) to position the bar cursor on the
protocol to change (No. 6 ‑ 15) and conrm the selection
with SELECT.
selecting the exercise test protocol to edit
configuration menu
Settings
Default Mode
Protocols
Contrast
Load Change
Language
Beep
Software Version
Date/Time
Training
Select
EKG Type
Choosefunction
Protocols
1. WHO
2. BAL
3. Hollmann
4. STD. France
5. Standard
6. 25W / 2min / 25W
7. 25W / 2min / 25W
8. 25W / 2min / 25W
9. 25W / 2min / 25W
Select
10. 25W / 2min / 25W
Choosefunction