Part 2 SES 2000 Vac (LS3) AUTOCLAVE
Page 12 of 53 ST-SM45g
PART 2 DESCRIPTION
ERROR CODE TABLE
Display Meaning / Cause
POWER FAILURE Temporary failure in the
(Error 1 not shown) mains supply to the unit.
Error 2 Faulty or incorrectly adjusted
door switch, or door not fully
closed at start of cycle.
Error 3 Water failed to enter chamber
from reservoir.
Error 4 Water level in chamber has
dropped during run-up to
sterilizing cycle.
Error 5 Heater element not working
during run-up to sterilizing phase.
Error 6 Control channel low temperature.
Error 7 Control channel high
temperature.
Error 8 Display channel low
temperature.
Error 9 Display channel high
temperature.
Error 10 Insufficient first vacuum pulse.
Error 11 Insufficient second vacuum
pulse.
Error 12 No steam pulse.
Error 13 (i) Before filling takes place
= Air Detector test failure
(ii) At beginning of sterilization =
Steam quality error.
Error 14 Insufficient drying vacuum.
Error 15 Sensor system failure.
Error 16 Clock speed error during
sterilizing phase.
Error 17 Band heater not achieving
setpoint temperature during
drying phase.
FILL RESERVOIR Water level in reservoir has
dropped below 'MIN' mark.
Overheating
30 In the unlikely event of overheating, the red overheat
warning lamp (see Fig. 2.1b and Fig. 2.4 item 8 ) at the
front of the autoclave will illuminate. If this happens, first
allow 10 to 15 minutes to elapse for the autoclave to cool,
then check the water level in the reservoir and top-up if
required. When the water level is correct, press the
'PRESS TO RESET' button at the rear of the cabinet (Fig.
2.4 item 21) and restart the cycle as normal. If the fault
persists, switch off the autoclave and call an Eschmann
Trained Engineer.