1. Electromagnetic emission
Guidance and manufacturer's declaration – electromagnetic emission
This device is intended for use in the electromagnetic environment specified below.
The customer or the end user of the device should assure that it is used in such an environment.
Emission test Compliance Electromagnetic environment - guidance
RF emission – CISPR 11 Group 1
This device uses RF energy only for its internal
function. Therefore its emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
Emission – CISPR 11 Class A
This device is suitable for use in all establishments
other than domestic and those directly connected to
the public low voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emission
IEC 61000-3-2
Class A
Voltage fluctuation/flicker emission IEC
61000-3-3
Complies
2. Electromagnetic immunity
Guidance and manufacturer's declaration – electromagnetic immunity
This device is intended for use in the electromagnetic environment specified below. The customer or the end
user of the ALM800 should assure that it is used in such an environment.
Immunity test
IEC 60601
Test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.
Transient/sequence of
rapid electrical pulses
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for I/O lines
±2 kV for power supply
lines
±1 kV for I/O lines
The quality of the network voltage
should be that of a typical commercial
or hospital environment.
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common mode
±1 kV differential mode
±2 kV common mode
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
<5 % U
T
(>95% dip in U
T
) for
0.5cycle
40 % U
T
(60% dip in U
T
) for 5
cycles
70 % U
T
(30% dip in U
T
) for 25
cycles
<5 % U
T
(>95% dip in U
T
) for
5s
<5 % U
T
(>95% dip in U
T
) for
0.5cycle
40 % U
T
(60% dip in U
T
) for 5
cycles
70 % U
T
(30% dip in U
T
) for 25
cycles
<5 % U
T
(>95% dip in U
T
) for 5s
The quality of the mains voltage
should be that of a typical commercial
or hospital environment. If the use of
the device requires continued
operation during mains power
outages, it is recommended to use
the device with an uninterruptible
power supply or batteries.
High frequency
(50/60Hz) magnetic
field IEC 61000-4-8
3 A/m 0.3A/m
If image distortion occurs, it may be
necessary to position the device
further from sources of power
frequency magnetic fields or to install
magnetic shielding. The power
frequency magnetic field should be
measured in the intended installation
location to assure that it is sufficiently
low.
Note: U
T
s the mains voltage in AC before applying the test level.
USER MANUAL > GUIDANCE AND MANUFACTURER'S DECLARATION
69 ALM800 - Auto lensmeter > V1 - 05-2020

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Essilor Instruments ALM 800 and is the answer not in the manual?

Essilor Instruments ALM 800 Specifications

General IconGeneral
BrandEssilor Instruments
ModelALM 800
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals