USER MANUAL > MAINTENANCE & SERVICING
Delta 2 - Edging system > v2 - 08.16 123
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
1.
2.
3.
2. Performing the edger autotest
This menu is used to carry out a self-diagnosis of the functions of the edger.
To test the proper performance of the edger, select > > > on your working screen.
Edger autotests
Locking of the door
Initialisation and rotation of lens shafts
Reproducer: lens shaft raising and lowering
Closing of lens shafts
Wheel rotation
GCD module rotation
Movement of the GCD module
Transfer initialization
Monoprobe test
Watering test
trough spraying
door spraying
Select the autotest to be carried out.
Close the door.
Press to start the test cycle.
To interrupt the cycle at any time, press .
> The result is displayed on the right of the icon of the autotest carried out:
The equipment is operational.
A malfunction has been detected, a description is displayed in the message box on the right.