USER MANUAL > AUTO-MAINTENANCE
TESS > V4 - 01.17 53
4
3
2
1
The functions described in this menu will allow you to correct issues yourself, or to direct the technician in solving any problems you
might encounter while using the tracer.
1. Calibration test
To ensure “high precision” performance is maintained, please carry out a calibration inspection regularly.
Press on to select the “calibration test” function.
Select the gauge caliber number on the tracer then, press on .
Position stably the frame gauge within the tracing boundary on the 4 lower studs.
The “V” must be open aiming up.
Press on ( ) of the tracer to launch tracing.
1. Test OK.
> The display indicates:
Press on and continue to use the tracer.
2. Test KO.
> The display indicates:
Press on and perform a complete calibration.
At the end of tracing, the clamps open, an audible signal indicates the end of calculations. Two possible cases:>